International Tables for Crystallography (2006). Vol. C, ch. 7.4, pp. 661-665

Section 7.4.4. White radiation and other sources of background

P. Suorttic

7.4.4. White radiation and other sources of background

| top | pdf |

7.4.4.1. Introduction

| top | pdf |

By definition, the background includes everything except the signal. In an X-ray diffraction measurement, the signal is the pattern of Bragg reflections. The profiles of the reflections should be determined by the structure of the sample, and so the broadening due to the instrument should be considered as background. In the ideal angle-dispersive experiment, a well collimated beam of X-rays having a well defined energy (and a polarization, perhaps) falls on the sample, and only the radiation scattered by the sample is detected. Furthermore, the detector should be able to resolve all the components of scattering by energy, so that each scattering process could be studied separately. It is obvious that only after this kind of analysis are the Bragg reflections (plus the possible disorder scattering) unequivocally separated from the background arising from other processes. In most cases, however, this analysis is not feasible, and the reflections are separated by using certain assumptions concerning their profile, and the success of this procedure depends on the peak-to-background ratio.

The ideal situation described above is all too often not encountered, and experimenters are satisfied with too low a level of resolution. The aim of the present article is to point out the sources of the unwanted and unresolved components of the registered radiation and to suggest how these may be eliminated or resolved, so that the quality of the diffraction pattern is as high as possible. The article can cover only a few of the possible experimental situations, and only the `almost ideal' angle-dispersive instrument is considered. It is assumed that the beam incident on the sample is monochromatized by reflection from a crystal and that the scattered radiation is registered by a low-noise quantum detector, which is the standard arrangement for modern diffractometers. Filtered radiation and photographic recording are used in certain applications, but these are excluded from the following discussion. The wavelength-dispersive or Laue methods are becoming popular at the synchrotron-radiation laboratories, and a short comment on these techniques will be included. Other sections of this volume deal with the components of scattering that are present even in the ideal experiment: thermal diffuse scattering (TDS), Compton and plasmon scattering, fluorescence and resonant Raman scattering, multiple scattering (coherent and incoherent), and disorder scattering.

The rest of the background may be termed `parasitic' scattering, and it arises from three sources:

  • (1) impurities of the incident beam;

  • (2) impurities of the sample;

  • (3) surroundings of the sample.

Parasitic scattering is occasionally mentioned in the literature, but it has hardly ever been the subject of a detailed study. Therefore, the present article will discuss the general principles of the minimization of the background and then illustrate these ideas with examples. Most of the discussion will be directed to the first of the three sources of parasitic scattering, because the other two depend on the details of the experiment.

7.4.4.2. Incident beam and sample

| top | pdf |

An ideal diffraction experiment should be viewed as an X-ray optical system where all the parts are properly matched for the desired resolution and efficiency. The impurities of the incident beam are the wavelengths and divergent rays that do not contribute to the signal but scatter from the sample through the various processes mentioned above. The propagation of the X-ray beam through the instrument is perhaps best illustrated by the so-called phase-space analysis. The three-dimensional version, which will be used in the following, was introduced by Matsushita & Kaminaga (1980a[link],b[link]) and was elaborated further by Matsushita & Hashizume (1983[link]). The width, divergence and wavelength range of the beam are given as a contour diagram, which originates in the X-ray source, and is modified by slits, monochromator, sample, and the detection system. The actual five-dimensional diagram is usually given as three-dimensional projections on the plane of diffraction and on the plane perpendicular to it and the beam axis, and in most cases the first projection is sufficient for an adequate description of the geometry of the experiment.

The limitations of the actual experiments are best studied through a comparison with the ideal situation. A close approximation to the ideal experimental arrangement is shown in Fig. 7.4.4.1[link] as a series of phase-space diagrams. The characteristic radiation from a conventional X-ray tube is almost uniformly distributed over the solid angle of 2π, and the relative width of the Kα1 or Kα2 emission line is typically Δλ/λ = 5 × 10−4. The acceptance and emittance windows of a flat perfect crystal are given in Fig. 7.4.4.1(b)[link]. The angular acceptance of the crystal (Darwin width) is typically less than 10−4 rad, and, if the width of the slit s or that of the crystal is small enough, none of the Kα2 distribution falls within the window. Therefore, it is sufficient to study the size and divergence distributions of the beam in the λ(Kα1) plane only, as shown in Fig. 7.4.4.1(c)[link]. The beam transmitted by the flat monochromator and a slit is shown as the hatched area, and the part reflected by a small crystal by the cross-hatched area. The reflectivity curve of the crystal is probed when the crystal is rotated. In this schematic case, almost 100% of the beam contributes to the signal. The typical reflection profile shown in Fig. 7.4.4.2[link] reveals the details of the crystallite distribution of the sample (Suortti, 1985[link]). The broken curve shows the calculated profile of the same reflection if the incident beam from a mosaic crystal monochromator had been used (see below).

[Figure 7.4.4.1]

Figure 7.4.4.1 | top | pdf |

Equatorial phase-space diagrams for a conventional X-ray source and parallel-beam geometry; x is the size and x′ = dx/dz the divergence of the X-rays. (a) Radiation distributions for two wavelengths, λ1 and λ2, at the source of width Δx, and downstream at a slit of width ±s1. (b) Acceptance and emittance windows of a flat perfect crystal, where the phase-space volume remains constant, AwaΔλ = EweΔλ, and the (x′, λ) section shows the reflection of a polychromatic beam (Laue diffraction). (c) Distributions for one wavelength at the source, flat perfect-crystal monochromator, sample (marked with the broken line), and the receiving slit (RS); z is the distance from the source.

[Figure 7.4.4.2]

Figure 7.4.4.2 | top | pdf |

Reflection 400 of LiH measured with a parallel beam of Mo Kα1 radiation (solid curve). The broken curve shows the reflection as convoluted by a Gaussian instrumentation function of 2σ = 0.1° and θ(α2) − θ(α1) = 0.13°, which values are comparable with those in Fig. 7.4.4.4[link].

The window of acceptance of a flat mosaic crystal is determined by the width of the mosaic distribution, which may be 100 times larger than the Darwin width of the reflection in question. This means that a convergent beam is reflected in the same way as from a bent perfect crystal in Johann or Johansson geometry. Usually, the window is wide enough to transmit an energy band that includes both [K\alpha_1] and [K\alpha_2] components of the incident beam. The distributions of these components are projected on the (x, x′, λ1) plane in Fig. 7.4.4.3[link]. The sample is placed in the (para)focus of the beam, and often the divergence of the beam is much larger than the width of the rocking curve of the sample crystal. This means that at any given time the signal comes from a small part of the beam, but the whole beam contributes to the background. The profile of the reflection is a convolution of the actual rocking curve with the divergence and wavelength distributions of the beam. The calculated profile in Fig. 7.4.4.2[link] demonstrates that in a typical case the profile is determined by the instrument, and the peak-to-background ratio is much worse than with a perfect-crystal monochromator.

[Figure 7.4.4.3]

Figure 7.4.4.3 | top | pdf |

Equatorial phase-space diagrams for two wavelengths, λ1 (solid lines) and λ2 (broken lines), projected on the plane λ = λ1. The monochromator at z = 200 mm is a flat mosaic crystal, and a small sample is located at z = 400 mm, as shown by the shaded area. The reflected beams at the receiving slit are shown for the (+, +) and (+, −) configurations of the monochromator and the sample.

An alternative arrangement, which has become quite popular in recent years, is one where the plane of diffraction at the monochromator is perpendicular to that at the sample. The beam is limited by slits only in the latter plane, and the wavelength varies in the perpendicular plane. An example of rocking curves measured by this kind of diffractometer is given in Fig. 7.4.4.4[link]. The [K\alpha_1] and [K\alpha_2] components are seen separately plus a long tail due to continuum radiation, and the profile is that of the divergence of the beam.

[Figure 7.4.4.4]

Figure 7.4.4.4 | top | pdf |

Two reflections of beryllium acetate measured with Mo Kα. The graphite (002) monochromator reflects in the vertical plane, while the crystal reflects in the horizontal plane. The equatorial divergence of the beam is 0.8°, FWHM.

In the Laue method, a well collimated beam of white radiation is reflected by a stationary crystal. The wavelength band reflected by a perfect crystal is indicated in Fig. 7.4.4.1(b)[link]. The mosaic blocks select a band of wavelengths from the incident beam and the wavelength deviation is related to the angular deviation by [\Delta\lambda/\lambda=\cot\theta\Delta\theta]. The angular resolution is determined by the divergences of the incident beam and the spatial resolution of the detector. The detector is not energy dispersive, so that the background arises from all scattering that reaches the detector. An estimate of the background level involves integrations over the incident spectrum at a fixed scattering angle, weighted by the cross sections of inelastic scattering and the attenuation factors. This calculation is very complicated, but at any rate the background level is far higher than that in a diffraction measurement with a monochromatic incident beam.

7.4.4.3. Detecting system

| top | pdf |

The detecting system is an integral part of the X-ray optics of a diffraction experiment, and it can be included in the phase-space diagrams. In single-crystal diffraction, the detecting system is usually a rectangular slit followed by a photon counter, and the slit is large enough to accept all the reflected beam. The slit can be stationary during the scan (ω scan) or follow the rotation of the sample (ω/2θ scan). The included TDS depends on these choices, but otherwise the amount of background is proportional to the area of the receiving slit. It is obvious from a comparison between Fig. 7.4.4.1[link] and Fig. 7.4.4.3[link] that a much smaller receiving slit is sufficient in the parallel-beam geometry than in the conventional divergent-beam geometry. Mathieson (1985[link]) has given a thorough analysis of various monochromator–sample–detector combinations and has suggested the use of a two-dimensional ω/2θ scan with a narrow receiving slit. This provides a deconvolution of the reflection profile measured with a divergent beam, but the same result with better intensity and resolution is obtained by the parallel-beam techniques.

The above discussion has concentrated on improving the signal-to-background ratio by optimization of the diffraction geometry. This ratio can be improved substantially by an energy-dispersive detector, but, on the other hand, all detectors have some noise, which increases the background. There have been marked developments in recent years, and traditional technology has been replaced by new constructions. Much of this work has been carried out in synchrotron-radiation laboratories (for references, see Thomlinson & Williams, 1984[link]; Brown & Lindau, 1986[link]).

A position-sensitive detector can replace the receiving slit when a reciprocal space is scanned. TV area detectors with an X-ray-to-visible light converter and two-dimensional CCD arrays have moderate resolution and efficiency, but they work in the current mode and do not provide pulse discrimination on the basis of the photon energy. One- and two-dimensional proportional chambers have a spatial resolution of the order of 0.1 mm, and the relative energy resolution, [\Delta E/E\simeq 0.2], is sufficient for rejection of some of the parasitic scattering.

The NaI(Tl) scintillation counter is used most frequently as the X-ray detector in crystallography. It has 100% efficiency for the commonly used wavelengths, and the energy resolution is comparable to that of a proportional counter. The detector has a long life, and the level of the low-energy noise can be reduced to about 0.1 counts s−1.

The Ge and Si(Li) solid-state detectors (SSD) have an energy resolution ΔE/E = 0.01 to 0.03 for the wavelengths used in crystallography. The relative Compton shift, Δλ/λ, is [(0.024\,{\rm \AA}/\lambda)\times(1-\cos2\theta)], where 2θ is the scattering angle, so that even this component can be eliminated to some extent by a SSD. These detectors have been bulky and expensive, but new constructions that are suitable for X-ray diffraction have become available recently. The effects of the detector resolution are shown schematically in Fig. 7.4.4.5[link] for a scintillation counter and a SSD.

[Figure 7.4.4.5]

Figure 7.4.4.5 | top | pdf |

Components of scattering at small scattering angles when the incident energy is just below the K absorption edge of the sample [upper part, (a)], and at large scattering angles when the incident energy is about twice the K-edge energy [upper part, (b)]. The abbreviations indicate resonant Raman scattering (RRS), plasmon (P) and Compton (C) scattering, coherent scattering (Coh) and sample fluorescence (Kα and Kβ). The lower part shows these components as convoluted by the resolution function of the detector: (a) a SSD and (b) a scintillation counter (Suortti, 1980b[link]).

Crystal monochromators placed in front of the detector eliminate all inelastic scattering but the TDS. The monochromator must be matched with the preceding X-ray optical system, the sample included, and therefore diffracted-beam monochromators are used in powder diffraction only (see Subsection 7.4.4.4[link]).

7.4.4.4. Powder diffraction

| top | pdf |

The signal-to-background ratio is much worse in powder diffraction than in single-crystal diffraction, because the background is proportional to the irradiated volume in both cases, but the powder reflection is distributed over a ring of which only the order of 1% is recorded. The phase-space diagrams of a typical measurement are shown in Fig. 7.4.4.6[link]. The Johansson monochromator is matched to the incident beam to provide maximum flux and good energy resolution. The Bragg–Brentano geometry is parafocusing, and, if the geometrical aberrations are ignored, the reflected beam is a convolution between the angular width of the monochromator focus (as seen from the sample) and the reflectivity curve of an average crystallite of the powder sample. The profile of this function is scanned by a narrow slit, as shown in the last diagram. The slit can be followed with a Johann or Johansson monochromator that has a narrow wavelength pass-band. In this case, there is no primary-beam monochromator, so that the incident beam at the sample is that given at z = 100 mm. The slit RS is the `source' for the monochromator, which focuses the beam at the detector. The obvious advantages of this arrangement are counterbalanced by certain limitations such as that the effective receiving slit is determined by the reflectivity curve of the monochromator, and this may vary over the effective area.

[Figure 7.4.4.6]

Figure 7.4.4.6 | top | pdf |

Equatorial phase-space diagrams for powder diffraction in the Bragg–Brentano geometry. (a) The acceptance and emittance windows of a Johannson monochromator; (b) the beam in the λ = λ1 plane: the exit beam from the Johansson monochromator is shown by the hatched area (z = 100 mm), the beam on the sample by two closely spaced lines, the reflectivity range of powder particles in a small area of the sample by the hatched area (z = 300 mm, note the change of scales), and the scan of the reflected beam by a slit RS by broken lines (z = 400 mm, at the parafocus).

Examples of a powder reflection measured with different instruments and 1.5 Å radiation are given in Fig. 7.4.4.7[link]. It should be noticed that scattering from the impurities of the sample and from the sample environment is negligible in all three cases. The width of the mosaic distribution of the 00.2 reflection of the pyrolytic graphite monochromator is 0.3°, which corresponds to a 180 eV (0.034 Å) wide transmitted beam. This is almost 10 times the separation between [K\alpha _1] and [K\alpha _2], and 70 times the natural width of the [K\alpha _1] line. The width of the focal line is about 0.2 mm, or 0.07°, and is seen as broadening of the reflection profile. The quartz (10.1) monochromator reflects a band that is determined by the projected width of the X-ray source. In the present case, the band is 15 eV wide, so that the monochromator can be tuned to transmit the [K\alpha_1] component only. The focal line is very sharp, 0.05 mm wide, and so the reflection is much narrower than in the preceding case. The third measurement was made with synchrotron radiation, and the receiving slit was replaced by a perfect-crystal analyser. The divergences of the incident and diffracted beams are about 0.1 mrad (less than 0.01°) in the plane of diffraction, so that the ideal parallel-beam geometry should prevail. However, the reflection is clearly broader than that measured with the conventional diffractometer. The reason is a wavelength gradient across the beam, which was monochromatized by a flat perfect crystal. On the other hand, the Ge (111) analyser crystal transmits elastic scattering and TDS only, and 2° away from the peak the background is 0.5% of the maximum intensity.

[Figure 7.4.4.7]

Figure 7.4.4.7 | top | pdf |

Three measurements of the 220 reflection of Ni at λ = 1.541 Å scaled to the same peak value; (a) in linear scale, (b) in logarithmic scale. Dotted curve: graphite (00.2) Johann monochromator, conventional 0.1 mm wide X-ray source (Suortti & Jennings, 1977[link]); solid curve: quartz (10.1) Johansson monochromator, conventional 0.05 mm wide X-ray source; broken curve: synchrotron radiation monochromatized by a (+, −) pair of Si (111) crystals, where the second crystal is sagittally bent for horizontal focusing (Suortti, Hastings & Cox, 1985[link]). The horizontal line indicates the half-maximum value. In all cases, the effective slit width is much less than the FWHM of the reflection.

The above considerations may seem to have little relevance to everyday crystallographic practices. Unfortunately, many standard methods yield diffraction patterns of poor quality. The quest for maximum integrated intensity has led to designs that make reflections broad and background high. It should be realized that not the flux but the brilliance of the incident beam is important in a diffraction measurement. The other aspect is that the information should not be lost in the experiment, and a divergent wide wavelength band is quite an ignorant probe of a reflection from a single crystal.

A situation where even small departures from the ideal diffraction geometry may cause large effects is measurement at an energy just below an absorption edge. Even a small tail of the energy band of the incident beam may excite radiation that becomes the dominant component of background. Similar effects are due to the harmonic energy bands reflected by most monochromators, particularly when the continuous spectrum of synchrotron radiation is used.

Scattering from the surroundings of the sample can be eliminated almost totally by shielding and beam tunnels. The general idea of the construction should be that an optical element of the instrument `sees' the preceding element only. Inevitably, the detector sees some of the environment of the sample. The density of air is about 1/1000 of that of a solid sample, so that 10 mm3 of irradiated air contributes to the background as much as a spherical crystal of 0.3 mm diameter. Strong spurious peaks may arise from slit edges and entrance windows of the specimen chamber, which should never be seen by the detector. A complete measurement without the sample is always a good starting point for an experiment.

References

Brown, G. S. & Lindau, I. (1986). Editors. Synchrotron radiation instrumentation. Proceedings of the International Conference on X-ray and VUV Synchrotron Radiation Instrumentation. Nucl. Instrum. Methods, A246, 511–595.
Mathieson, A. McL. (1985). Small-crystal X-ray diffractometry with a crystal ante-monochromator. Acta Cryst. A41, 309–316.
Matsushita, T. & Hashizume, H. (1983). Handbook of synchrotron radiation, Vol. I, edited by E. E. Koch, pp. 261–314. Amsterdam: North-Holland.
Matsushita, T. & Kaminaga, U. (1980a). A systematic method of estimating the performance of X-ray optical systems for synchrotron radiation. I. Description of various optical elements in position-angle space for ideally monochromatic X-rays. J. Appl. Cryst. 13, 465–471.
Matsushita, T. & Kaminaga, U. (1980b). A systematic method of estimating the performance of X-ray optical systems for synchrotron radiation. II. Treatment in position-angle-wavelength space. J. Appl. Cryst. 13, 472–478.
Suortti, P. (1980b). Components of the total X-ray scattering. Accuracy in powder diffraction, edited by S. Block & C. R. Hubbard, pp. 1–20. Natl Bur. Stand. (US) Spec. Publ. No. 567.
Suortti, P. (1985). Parallel beam geometry for single-crystal diffraction. J. Appl. Cryst. 18, 272–274.
Suortti, P., Hastings, J. B. & Cox, D. E. (1985). Powder diffraction with synchrotron radiation. I. Absolute measurements. Acta Cryst. A41, 413–416.
Suortti, P. & Jennings, L. D. (1977). Accuracy of structure factors from X-ray powder intensity measurements. Acta Cryst. A33, 1012–1027.