Figure 2.1.10 

A. Kerna*

aBruker AXS, Östliche Rheinbrückenstrasse 49, Karlsruhe 76187, Germany
Correspondence e-mail: arnt.kern@bruker-axs.de

[Figure 2.1.10]
Figure 2.1.10

Example of counterbalancing of a vertical θ–θ goniometer. The counterweights (grey parts) are located at positions matching the weights and locations of the X-ray source and detector. Mounting of different beam-path components with significantly different weight or moving of, for example, the X-ray source and/or the detector to change the respective radii may require repositioning of the counterweights to maintain goniometer accuracy and instrument alignment.