2.5.4.1. Phase identification

| top | pdf |

In materials science, a phase is defined as a region that has uniform chemical composition and physical properties, including crystal structure. Therefore, every phase should give a unique diffraction pattern. A sample for X-ray diffraction may contain a single phase or multiple phases. Analysis of the diffraction pattern can accurately and precisely determine the contents of the sample. This qualitative analysis is called phase identification (phase ID). One of the most efficient methods of phase identification is to compare the diffraction pattern from an unknown material to those in a database of a large number of standard diffraction patterns. The most comprehensive database is the Powder Diffraction File (PDF), updated annually by the International Centre for Diffraction Data (ICDD).

Two-dimensional X-ray diffraction has enhanced phase identification in many respects (Rudolf & Landes, 1994[link]; Sulyanov et al., 1994[link]; Hinrichsen, 2007[link]). Because of its ability to collect diffracted X-rays in a large angular range in both the 2θ and γ directions, it can collect diffraction data with high speed and better sampling statistics than obtained by conventional diffraction. Owing to point-beam illumination on the sample, a relatively small sample size is required for phase identification. The large 2D detector allows for a large 2θ range to be analysed without any movement of the sample and detector. This makes it possible to perform in situ phase investigation on samples during phase transformations, chemical reactions and deformations. The diffraction information in the γ direction allows accurate phase identification of samples with large grains and preferred orientation.

In the Bragg–Brentano geometry, the 2θ resolution is controlled by the selection of the divergence slit and receiving slit in the diffractometer plane, and the axial divergence is controlled by Soller slits, while in a diffractometer with a 2D detector, the 2θ resolution is mainly determined by the spatial resolution of the detector and the sample-to-detector distance. The relative peak intensity in a diffraction pattern from a sample with texture measured with a 2D detector can be significantly different from the results measured with Bragg–Brentano geometry. It is imperative to study the nature of these discrepancies so that the diffraction patterns collected with 2D detectors can be used for phase ID with proper understanding and correction if necessary.

When two-dimensional diffraction is used for phase identification, the first step is to integrate the 2D diffraction frame into a diffraction profile resembling the diffraction pattern collected with a conventional diffractometer (Cervellino et al., 2006[link]; Rodriguez-Navarro, 2006[link]; Boesecke, 2007[link]; Fuentes-Montero et al., 2011[link]; Hammersley, 2016[link]). The integrated diffraction profiles can be analysed with all existing algorithms and methods, including profile fitting with conventional peak shapes and fundamental parameters, quantification of phases, and lattice-parameter determination and refinement (Ning & Flemming, 2005[link]; Flemming, 2007[link]; Jabeen et al., 2011[link]). The results can be used to search a powder-diffraction database to find possible matches. Since there is a great deal of literature covering these topics (Cullity, 1978[link]; Jenkins & Snyder, 1996[link]; Pecharsky & Zavalij, 2003[link]), this section will focus on the special characteristics of two-dimensional X-ray diffraction as well as system geometry, data-collection strategies and data analysis in dealing with relative peak intensities, 2θ resolution, grain size and distribution, and preferred orientation. Many factors and correction algorithms described here can help in understanding the characteristics of two-dimensional diffraction. In most applications, however, the γ-integrated profile can be used for phase identification without these corrections.

2.5.4.1.1. Relative intensity

| top | pdf |

The integrated intensity diffracted from polycrystalline materials with a random orientation distribution is given by[{{{I}}_{hkl}} = {k_I}{{{p_{hkl}}} \over {{v^2}}}({\rm LPA}){\lambda ^3}F_{hkl}^2{g_{hkl}}(\alpha, \beta)\exp \left({ - 2{M_t} - 2{M_s}} \right),\eqno(2.5.45)]where kI is an instrument constant that is a scaling factor between the experimental observed intensities and the calculated intensity, phkl is the multiplicity factor of the crystal plane (hkl), v is the volume of the unit cell, (LPA) is the Lorentz–polarization and absorption factors, λ is the X-ray wavelength, [F_{hkl}] is the structure factor for the crystal plane (hkl), [{g_{hkl}}(\alpha, \beta)] is the normalized pole-density distribution function and exp(−2Mt − 2Ms) is the attenuation factor due to lattice thermal vibrations and weak static displacements (Warren, 1990[link]; He et al., 1994[link]). Except for the texture effect, all the factors in the above equation are either discussed in the previous sections or have the same definitions and values as in conventional diffraction.

Phase-identification studies by XRD are preferably carried out on powders or polycrystalline samples with a random orientation distribution of crystallites. Preferred orientation causes relative intensities to deviate from theoretical calculations or those reported in reference databases. In practice, a sample with a perfectly random orientation distribution of crystallites is very hard to fabricate and most polycrystalline samples have a preferred orientation to a certain extent. Discrepancies in the relative peak intensities between conventional diffraction and 2D-XRD are largely due to texture effects. For B-B geometry, the diffraction vector is always perpendicular to the sample surface. With a strong texture, it is possible that the pole density of certain reflections in the sample normal direction is very low or even approaches zero. In this case, the peak does not appear in the diffraction pattern collected in B-B geometry. In 2D-XRD, several diffraction rings may be measured with a single incident beam; the corresponding diffraction vectors are not necessarily in the sample normal direction. The diffraction profiles from 2D frames are produced by γ integration, therefore the texture factor [{g_{hkl}}(\alpha, \beta)] should be replaced by the average normalized pole-density function within the γ integration range [\left\langle {{g_{hkl}}(\Delta \gamma)} \right\rangle ]. The relation between (α, β) and (2θ, γ) is given in Chapter 5.4[link] . The chance of having zero pole density over the entire γ-integration range is extremely small. Therefore, phase identification with 2D-XRD is much more reliable than with conventional diffraction.

2.5.4.1.2. Detector distance and resolution

| top | pdf |

The 2θ resolution with B-B geometry is controlled by the size of the slits. Smaller apertures of the divergence slit are used for higher 2θ resolution and larger apertures for fast data collection. With a two-dimensional X-ray diffraction system, the 2θ resolution is achieved with different approaches. A flat 2D detector has the flexibility to be used at different sample-to-detector distances. The detector resolution is determined by the pixel size and point-spread function. For the same detector resolution and detector active area, a higher resolution can be achieved at larger distance, and higher angular coverage at shorter distance. The sample-to-detector distance should be optimized depending on the 2θ measurement range and required resolution. In situations where the 2θ range of one frame is not enough, several frames at sequential 2θ ranges can be collected. The integrated profiles can then be merged to achieve a large 2θ range. Fig. 2.5.16[link] shows four 2D frames collected from a battery material with a microgap detector. The slice integration region is defined by two conic lines and two horizontal lines. The diffraction profile integrated from the merged frames is displayed below.

[Figure 2.5.16]

Figure 2.5.16 | top | pdf |

Diffraction pattern merged from four 2D frames collected from a battery material.

2.5.4.1.3. Defocusing effect

| top | pdf |

A 2D diffraction pattern over a range of 2θ is measured simultaneously with a single incident angle, so the incident angle has to be lower than the minimum 2θ angle. Since the reflected angle cannot always be the same as the incident angle, geometric aberrations are observed. The defocusing effect occurs when the incident angle is lower than the reflection angle. At low incident angles, the incident beam spreads over the sample surface into an area much larger than the size of the original X-ray beam. The observed diffracted beam size is magnified by the defocusing effect if the diffracted beam makes an angle larger than the incident angle. The defocusing effect for reflection-mode diffraction can be expressed as[{{B} \over {b}} = {{\sin \theta _2} \over {\sin \theta _1}} = {{\sin (2\theta - \omega)} \over {\sin \omega }},\eqno(2.5.46)]where θ1 is the incident angle, b is the incident beam size and B is diffracted beam size. The ratio of B to b is a measurement of the geometric aberration and will be referred to as the defocusing factor. In principle, defocusing occurs only when B/b is larger than 1. The reflected beam is actually focused to the detector when θ2 < θ1. The defocusing effect occurs when θ2 > θ1 and the defocusing factor increases with increasing θ2 or decreasing θ1. The maximum defocusing appears at θ2 = 90°. For the θ–2θ configuration, the incident angle ω (= θ1) is used in the equation.

For B-B geometry with a divergent slit and receiving slit of the same size the defocusing factor is always 1. With a 2D detector the defocusing factor varies with the 2θ angle. If a large 2θ range is measured on a flat sample in reflection mode, it is always desirable to collect several frames at different incident angles for each 2θ range so as to improve the 2θ resolution. A cylindrical detector may collect a diffraction pattern over a large 2θ range (Gelfi et al., 2005[link]). However, the defocusing effect prevents it from being used for a large 2θ range for a flat sample. Fig. 2.5.17[link] compares the effect for a flat detector and a cylindrical detector. Fig. 2.5.17[link](a) shows a cylindrical detector being used to collect a diffraction pattern from a flat sample for a 2θ range of 5 to 80°. The incident angle must be kept at 5° or lower. Fig. 2.5.17[link](b) shows a flat detector being used to collect the diffraction pattern over the same 2θ range. In order to minimize the defocusing effect, the data collection is done at four different incident angles (5, 15, 25 and 35°) with four corresponding detector swing angles (10, 30, 50 and 70°). Fig. 2.5.17[link](c) compares the defocusing factors of the two configurations. The horizontal dot-dashed line with defocusing factor B/b = 1 represents the situation with B-B geometry. The defocusing factor continues to increase with 2θ angle up to B/b = 11 for cylindrical detector. That means that the 2θ resolution would be 10 times worse than for the B-B geometry. For the diffraction pattern collected with a flat detector in four steps, the defocusing factor fluctuates above 1, with the worst value being less than 3. Another approach to avoiding defocusing is to collect the diffraction pattern in transmission mode. There is no defocusing effect in transmission when the incident beam is perpendicular to the sample surface. Therefore, the transmission pattern has significantly better 2θ resolution. Transmission-mode diffraction also has other advantages. For instance, the air scattering from the primary beam may be blocked by a flat sample, therefore lowering the background from air scattering. However, transmission-mode diffraction data can only be collected from samples with limited thickness, and the maximum scattering intensity is achieved at low 2θ angles with a sample thickness of [t = 1/\mu ], where μ is the linear absorption coefficient. The scattering intensity drops dramatically when the thickness increases.

[Figure 2.5.17]

Figure 2.5.17 | top | pdf |

Defocusing effects: (a) cylindrical detector; (b) flat detector at various incident angles and detector swing angles; (c) comparison of defocusing factors.

2.5.4.1.4. Sampling statistics

| top | pdf |

In powder X-ray diffraction, the number of crystallites contributing to each reflection must be sufficiently large to generate reproducible integrated peak intensities (see Chapter 2.10[link] ). A larger number of contributing crystallites gives better precision or sampling statistics (also referred to as particle statistics). Sampling statistics are determined by both the structure of the sample and the instrumentation. For a powder sample in which the crystallites are perfectly randomly oriented, the number of contributing crystallites for a diffraction peak can be given as[N_s = p_{hkl} {Vf_i \over v_i} {\Omega \over 4\pi },\eqno(2.5.47)]where phkl is the multiplicity of the diffracting planes, V is the effective sampling volume, fi is the volume fraction of the measuring crystallites (fi = 1 for single-phase materials), vi is the volume of individual crystallites and Ω is the angular window of the instrument (given as a solid angle). The multiplicity term, phkl, effectively increases the number of crystallites contributing to the integrated intensity from a particular set of (hkl) planes. The volume of individual crystallites, vi, is an average of various crystallite sizes. The combination of the effective sampling volume and the angular window makes up the instrumental window, which determines the total volume of polycrystalline material making a contribution to a Bragg reflection. For 2D-XRD, the instrumental window is not only determined by the incident beam size and divergence, but also by the detective area and the sample-to-detector distance (γ angular coverage).

In B-B geometry, the effective irradiated volume is a constant,[V_{\rm BB} = A_oA_{\rm BB} = {A_o/2\mu },\eqno(2.5.48)]where Ao is the cross-section area of the incident beam measured on the sample surface, [A_{\rm BB} = 1/(2\mu)] is the transmission coefficient for B-B geometry, and μ is the linear absorption coefficient. For 2D-XRD, the effective volume is given as[V = A_oA = A_oT/2\mu,\eqno(2.5.49)]where A is the transmission coefficient and T is the transmission coefficient with B-B normalization for either transmission or reflection as given previously.

The angular window is given as a solid angle. The incident beam has a divergence angle of β1 within the diffraction plane and β2 in the perpendicular direction. The angular window corresponding to the incident-beam divergence is given by[\Omega = {\beta _1}{\beta _2}/\sin \theta \ {\rm or}\ \Omega = {\beta ^2}/\sin \theta \ {\rm if}\ \beta = {\beta _1} = {\beta _2}.\eqno(2.5.50)]

For 2D-XRD, the angular window is not only determined by the incident-beam divergence, but also significantly increased by γ integration. When γ integration is used to generate the diffraction profile, it actually integrates the data collected over a range of various diffraction vectors. Since the effect of γ integration on sampling statistics is equivalent to the angular oscillation on the ψ axis in a conventional diffractometer, the effect is referred to as virtual oscillation and Δψ is the virtual oscillation angle. In conventional oscillation, mechanical movement may result in some sample-position error. Since there is no actual physical movement of the sample stage during data collection, virtual oscillation can avoid this error. This is crucial for microdiffraction. The angular window with the contributions of both the incident-beam divergence and the virtual oscillation is[\Omega = \beta \Delta \psi = 2\beta \arcsin [\cos \theta \sin (\Delta \gamma /2)],\eqno(2.5.51)]where β is the divergence of the incident beam. While increasing the divergence angle β may introduce instrumental broadening which deteriorates the 2θ resolution, virtual oscillation improves sampling statistics without introducing instrumental broadening.

In the cases of materials with a large grain size or preferred orientation, or of microdiffraction with a small X-ray beam size, it can be difficult to determine the 2θ position because of poor counting statistics. In these cases, some kind of sample oscillation, either by translation or rotation, can bring more crystallites into the diffraction condition. Angular oscillation is an enhancement to the angular window of the instrument. The effect is that the angular window scans over the oscillation angle. Any of the three rotation angles (ω, ψ, ϕ) or their combinations can be used as oscillation angles. Angular oscillation can effectively improve the sampling statistics for both large grain size and preferred orientation. As an extreme example, a powder-diffraction pattern can be generated from single-crystal sample if a sufficient angular window can be achieved by sample rotation in such as way as to simulate a Gandolfi camera (Guggenheim, 2005[link]). Sample oscillation is not always necessary if virtual oscillation can achieve sufficient sampling statistics.

References

Boesecke, P. (2007). Reduction of two-dimensional small- and wide-angle X-ray scattering data. J. Appl. Cryst. 40, s423–s427.Google Scholar
Cervellino, A., Giannini, C., Guagliardi, A. & Ladisa, M. (2006). Folding a two-dimensional powder diffraction image into a one-dimensional scan: a new procedure. J. Appl. Cryst. 39, 745–748.Google Scholar
Cullity, B. D. (1978). Elements of X-ray Diffraction, 2nd ed. Reading, MA: Addison-Wesley.Google Scholar
Flemming, R. L. (2007). Micro X-ray diffraction (µXRD): a versatile technique for characterization of Earth and planetary materials. Can. J. Earth Sci. 44, 1333–1346.Google Scholar
Fuentes-Montero, L., Montero-Cabrera, M. E. & Fuentes-Cobas, L. (2011). The software package ANAELU for X-ray diffraction analysis using two-dimensional patterns. J. Appl. Cryst. 44, 241–246.Google Scholar
Gelfi, M., La Vecchia, G. M., Lecis, N. & Troglio, S. (2005). Relationship between through-thickness residual stress of CrN-PVD coatings and fatigue nucleation sites. Surf. Coat. Technol. 192, 263–268.Google Scholar
Guggenheim, S. (2005). Simulations of Debye–Scherrer and Gandolfi patterns using a Bruker Smart Apex diffractometer system. Bruker AXS Application Note 373.Google Scholar
Hammersley, A. P. (2016). FIT2D: a multi-purpose data reduction, analysis and visualization program. J. Appl. Cryst. 49, 646–652.Google Scholar
He, B., Rao, S. & Houska, C. R. (1994). A simplified procedure for obtaining relative x-ray intensities when a texture and atomic displacements are present. J. Appl. Phys. 75, 4456–4464.Google Scholar
Hinrichsen, B. (2007). Two-dimensional X-ray powder diffraction. PhD dissertation, University of Stuttgart, Germany.Google Scholar
Jabeen, S., Raza, S. M., Ahmed, M. A., Zai, M. Y. & Erlacher, K. (2011). Two dimensional X-ray diffraction (2D-XRD) studies on olivine of U. S. A. J. Chem. Soc. Pak. 33(5), 612–618.Google Scholar
Jenkins, R. & Snyder, R. L. (1996). Introduction to X-ray Powder Diffractometry. New York: John Wiley & Sons.Google Scholar
Ning, G. & Flemming, R. L. (2005). Rietveld refinement of LaB6: data from µXRD. J. Appl. Cryst. 38, 757–759.Google Scholar
Pecharsky, V. K. & Zavalij, P. Y. (2003). Fundamentals of Powder Diffraction and Structure Characterization of Materials. Boston: Kluwer Academic Publishers.Google Scholar
Rodriguez-Navarro, A. B. (2006). XRD2DScan: new software for polycrystalline materials characterization using two-dimensional X-ray diffraction. J. Appl. Cryst. 39, 905–909.Google Scholar
Rudolf, P. R. & Landes, B. G. (1994). Two-dimensional X-ray diffraction and scattering of microcrystalline and polymeric materials. Spectroscopy, 9(6), 22–33.Google Scholar
Sulyanov, S. N., Popov, A. N. & Kheiker, D. M. (1994). Using a two-dimensional detector for X-ray powder diffractometry. J. Appl. Cryst. 27, 934–942.Google Scholar