Section 2.5.4.1.2. Detector distance and resolution

B. B. Hea*

aBruker AXS Inc., 5465 E. Cheryl Parkway, Madison, WI 53711, USA
Correspondence e-mail: bob.he@bruker.com

2.5.4.1.2. Detector distance and resolution

| top | pdf |

The 2θ resolution with B-B geometry is controlled by the size of the slits. Smaller apertures of the divergence slit are used for higher 2θ resolution and larger apertures for fast data collection. With a two-dimensional X-ray diffraction system, the 2θ resolution is achieved with different approaches. A flat 2D detector has the flexibility to be used at different sample-to-detector distances. The detector resolution is determined by the pixel size and point-spread function. For the same detector resolution and detector active area, a higher resolution can be achieved at larger distance, and higher angular coverage at shorter distance. The sample-to-detector distance should be optimized depending on the 2θ measurement range and required resolution. In situations where the 2θ range of one frame is not enough, several frames at sequential 2θ ranges can be collected. The integrated profiles can then be merged to achieve a large 2θ range. Fig. 2.5.16[link] shows four 2D frames collected from a battery material with a microgap detector. The slice integration region is defined by two conic lines and two horizontal lines. The diffraction profile integrated from the merged frames is displayed below.

[Figure 2.5.16]

Figure 2.5.16 | top | pdf |

Diffraction pattern merged from four 2D frames collected from a battery material.








































to end of page
to top of page